DEVELOPMENT OF VALIDATED STABILITY INDICATING RPHPLC METHOD FOR SIMULTANEOUS ESTIMATION OF AMLODIPINE BESYLATE AND CANDESARTAN CILEXETIL FROM TABLET
Y. F. Padole, S. K. Rabade, P. S. Jirvankar, M. J. Umekar and R. T. Lohiya*
ABSTRACT
A simple, isocratic, rapid and accurate reversed phase high
performance liquid chromatography method was developed for the
quantitative determination of Candesartan Cilexetil and Amlodipine
Besylate tablets. The chromatographic separation was achieved on
Water Xterra R18, 150×4.6 mm, 3.5u (C18) using Mobile Phase A :
ACN: Water: OPA (950:50:01) and Mobile Phase B: ACN: Water:
OPA (50:950:01), and the λmax of Amlodipine Besylate was detected
at 237nm, where Candesartan Cilexetil exhibits sufficient absorbance
at 254 nm. The linear range for Candesartan Cilexetil and Amlodipine
Besylate were (2.8-42ppm) and (6.4-96) was obtained with correlation
coefficients ≥0.999 for each analyte. The retention time were found to
be 4.2 and 8.5 min Candesartan Cilexetil and Amlodipine Besylate respectively. Candesartan
Cilexetil and Amlodipine Besylate was subjected to stress conditions (hydrolysis (acid, base)
oxidation, photolysis, thermal degradation and humidity degradation) and the stressed
samples were analyzed by use of the method. The major degradation was observed in base
and minor in acid, thermal, oxidation, humidity and photolysis. The forced degradation
studies prove the stability indicating power of the method.
Keywords: Amlodipine Besylate (AMLO) and Candesartan Cilexetil (CANDE), Method validation, RP- HPLC, C18, Degradation Studies.
[Full Text Article]