WJPR Citation
|
| All | Since 2020 | |
| Citation | 8502 | 4519 |
| h-index | 30 | 23 |
| i10-index | 227 | 96 |
DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE DETERMINATION OF SERTRALINE IN BULK DRUG AND DOSAGE FORM
Shriram H. Bairagi* and Dr. R. S. Ghosh
Abstract The present study indicates a simple, accurate and precise RP-HPLC method for the estimation of Sertraline in bulk and in dosage Form. The mobile phase used was ACN: Phosphate buffer = 80:20 (pH 2.9). The specification of the chromatographic system is waters C18, 5m, 150mm x4.6mm, flow rate 1.0ml/min, detection 205nm, and injection volume 20μl and run time 10 min. The retention times found to be 2.30 min. The linearity was in the range of 5-50 μg/ml. The recoveries found to be 99.25 to 100.62%. The proposed method was validated and successfully applied to the estimation of Sertraline and its dosage forms. Keywords: RP-HPLC, Sertraline, Analytical Method development, Anti-Depressant activity. [Full Text Article] [Download Certificate] |