WJPR Citation
|
| All | Since 2020 | |
| Citation | 8502 | 4519 |
| h-index | 30 | 23 |
| i10-index | 227 | 96 |
A NOVEL ANALYTICAL METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF CAPMATINIB IN BULK AND MARKETED FORMULATION BY RP-HPLC
Valluri Naga Sravani*, Kandhagatla Saisneha, Pragathi Talusani and Rajani Lagisetty
Abstract A new, simple, rapid, precise, accurate and reproducible RP-HPLC method for estimation of Capmatinib in bulk form and marketed formulation. Separation of Capmatinib was successfully achieved on a Develosil ODS HG-5 RP C18, 5μm, 15cmx4.6mm i.d. column in an isocratic mode of separation utilizing Methanol : Phosphate buffer (0.02M, pH-3.6) in the ratio of 45:55% v/v at a flow rate of 1.0 mL/min and the detection was carried out at 267nm. The method was validated according to ICH guidelines for linearity, sensitivity, accuracy, precision, specificity and robustness. The response was found to be linear in the drug concentration range of 12-28mcg/mL for Capmatinib. The correlation coefficient was found to be 0.9995 for Capmatinib. The LOD and LOQ for Capmatinib were found to be 5.004μg/mL and 15.164μg/mL respectively. The proposed method was found to be good percentage recovery for Capmatinib, which indicates that the proposed method is highly accurate. The specificity of the method shows good correlation between retention times of standard solution with the sample solution. Therefore, the proposed method specifically determines the analyte in the sample without interference from excipients of pharmaceutical dosage forms. Keywords: Capmatinib, RP-HPLC, Accuracy, Precision, Robustness, ICH Guidelines. [Full Text Article] [Download Certificate] |