SIMULTANEOUS ESTIMATION OF METHOTREXATE AND TRETINOIN BY REVERSE PHASE –HPLC METHOD
Aditya S., Arun P.*, Neelesh D., Neeraj K., Shailendra P., Shivani T. and Stuti P.
Abstract
A simple fast, precise and accurate liquid chromatographic method was
developed for simultaneous estimation of Methotrexate(MTX) and
tretioin(ATX) combined dosage form by RP-HPLC (Reverse Phase
High Performance Liquid Chromatography) method. Chromatography
was out on a column -18 and ACN:Buffer (85:15 v/v) of resultant pH
3.0, as mobile phase, at a flow rate of 1.2ml/min and measured at
340nmth. The calibration curves linearity range were 1-6ug/ml for
methorexate and 1-16 ug/ml for tretioin, regression coefficient values
are 0.9989 and 0.9999 respectively. The average retention time of
Methotrexate and trition were found to be 1.95 and 8.98 min
respectively. The method was reproducible, with good resolution and sharp peak of
Methotrexate and tretioin. There results of the analysis have been validated statistically.
Keywords: METHOTREXATE, TRETIONIN, RP-HPLC, ASSAY, UV DETECTION.